CADD Legacy Plus 6500 Operator's Manual page 41

Ambulatory infusion pump
Table of Contents

Advertisement

• Press
.
11. Verify the Upstream Sensor status.
• Make sure the desired setting is displayed. This screen will show
whether the Upstream Occlusion Sensor is turned on or off.
WARNING: When the Upstream Occlusion Sensor is turned off,
the pump will not detect occlusions upstream (between pump and
fluid container). Periodically inspect the fluid container for de-
creasing volume, inspect the fluid path for kinks, a closed clamp,
or other upstream occlusions. Upstream occlusions could result in
under- or non-delivery of medications. If undetected, these occlu-
sions could result in death or serious injury to the patient.
• If you need to change the Upstream Occlusion Sensor setting, see
Section 4, Biomed Functions.
• Press
.
12. Review the program.
Press
repeatedly to review the programming screens. If you
need to reprogram a setting, press
screen appears and change the setting as described in this section.
Section 2: Pump Setup and Programming