Section 5: Reference
Messages and Alarms
High Pressure
T
-
wo
Tone alarm
Key pressed,
Please release
T
-
wo
Tone alarm
LowBat
T
-
hree Two
Tone beeps
5
every
minuTes
Motor Locked,
remove all power
T
-
wo
Tone alarm
[No message]
T
-
wo
Tone alarm
0
Description/Corrective Action
The pump has detected high pressure,
which may be resulting from a
downstream blockage, kink in the fluid
path, or a closed tubing clamp. Remove
the obstruction to resume operation.
„
⁄
Or, press
or
and silence the alarm for 2 minutes, then
remove the obstruction and restart the
pump.
If a key is being pressed, stop pressing
it. If the alarm persists, close the tubing
clamp and remove the pump from use.
Contact Customer Service to return the
pump for service.
The batteries are low, but the pump is
still operable.
• Change the batteries soon.
Batteries are depleted and the pump was
powered up with the AC Adapter. Install
new AA batteries, reconnect the AC
adapter, and restart the pump.
With no AC adapter attached, the batter-
ies have been removed while the pump is
running. The pump is now stopped and
unpowered. Install batteries to silence the
alarm.
OR
Batteries were removed within
approximately 15 seconds after stopping the
pump. Install new batteries to silence the
alarm, if desired. Otherwise, the alarm will
stop within a short period of time.