Section 2: Pump Setup and Programming

Programming the Pump: General Instructions

The procedure for changing a programmed setting is similar for most
programming screens.
WARNING: Ensure that the ± 6% System Delivery Accuracy specifica-
tion is taken into account when programming the pump and/or filling
the Medication Cassette Reservoir. Failure to do so may result in
medication in the reservoir becoming depleted sooner than expected.
If the pump is being used to deliver critical or life sustaining medica-
tion, the interruption in the delivery of medication could result in
patient injury or death.
Make sure the pump is stopped and in Lock Level 0.
To begin programming, start at the main screen and press
To change a setting, press
pears. (Press and hold these keys to change values with increasing
speed.)
Press
¤
within 25 seconds to confirm a change or the screen will
revert to the previous setting.
If any key other than
Press
to return to the screen being programmed, scroll to the
desired value, and press
Press
to advance to the next screen.
To leave a setting unchanged, press

´
or
Î
until the desired setting ap-
¤
is pressed, "Value not saved" will appear.
¤
.
to go to the next screen.
.