Figure 4-8: Waveform Advance Mode Initial Test Hookup - Sony Tektronix AWG2021 Service Manual

Arbitrary waveform generator
Table of Contents

Advertisement

Check Waveform Advance
Mode
AWG2021 Service Manual
Electrical
Operating mode, Waveform Advance, on page 1-4.
Characteristic
Checked
Equipment
One 50 W coaxial cable
Required
One oscilloscope
Prerequisites
The AWG2021 meets the prerequisites listed on page 4-7.
1. Install test hookup and set test equipment controls:
a.
Hook up the oscilloscope: Connect the AWG2021 CH1 output
connector through the coaxial cable to the CH1 vertical input connector
on the oscilloscope (see Figure 4–8).
AWG2021
AWG2021

Figure 4-8: Waveform Advance Mode Initial Test Hookup

b. Set oscilloscope controls:
Vertical
. . . . . . . . . . . . . . . . . . . . . .
CH1 coupling
CH1
. . . . . . . . . . . . . . . . . . . . .
CH1 input impedance
Horizontal
Sweep
. . . . . . . . . . . . . . . . . . . .
Trigger
Source
. . . . . . . . . . . . . . . . . . .
Coupling
. . . . . . . . . . . . . . . . . .
Slope
. . . . . . . . . . . . . . . . . . . .
Level
. . . . . . . . . . . . . . . . . . . .
Mode
. . . . . . . . . . . . . . . . . . . .
2. Set the AWG2021 controls and select the waveform file:
a. Initialize AWG2021 controls: Push UTILITY!Misc!Config...!Re-
set to Factory!O.K.
Oscilloscope
TDS
540
CH1
. . . . . . . . . . . . . .
DC
0.2 V/div.
50 W
. . . . . . .
5 ms/div.
CH1
DC
Positive
0 V
Auto
Performance Tests
4-21