Removal and Installation Procedures

Rear BNC Connector

6-34
1
d. Using a
/
-inch nut driver, remove the IEEE STD 488 PORT cover by
4
removing the two spacer posts.
e. Grasp the serial interface connector and pull the rear shield cover toward
you taking care not to damage the cable. Then remove the rear shield
cover.
f.
You can remove the rear panel of a slot in which no board is mounted.
4. Remove serial interface connector: Use a 5 mm nut driver to remove the nut
attaching the connector to the rear shield cover; then, remove the connector.
5. Reinstallation: Install the serial interface connector if you removed it by
reversing the items in step 4; then, install the rear shield cover by doing
substeps 3a–3f in reverse order.
1. Assemble equipment and locate modules to be removed:
a. You will need a
b. Locate the modules to be removed in the locator diagram, Internal
Modules, in Figure 6–4. (An instrument with Option 02 has two
additional BNC connectors.)
2. Orient instrument: Set the AWG2021 with the top down on the work surface
and the left side facing you (see Figure 6–17).
3. Remove rear BNC connector:
a. Disconnect the interconnect cable at the AM IN connector.
b. Using the open-end wrench, remove the nut securing the BNC connector
to the chassis. Then, lift the BNC connector out of the chassis to
complete the removal.
c. For the instrument with Option 02, remove the CH2 MARKER1 OUT
and CH2 SYNC OUT connectors using the same procedure.
4. Reinstallation: Install the rear BNC connector by doing substeps 3a–3c in
reverse order.
@
1
-inch open-end wrench to do this procedure.
2