Performance Tests
Check Burst Mode
4-16
5. Check triggered mode with external triggering:
a. Enable function generator output: Turn on the function generator output.
b. Check triggering: Check that for each trigger supplied by the function
generator, the oscilloscope displays a one-cycle sine wave.
6. End procedure: Turn off the function generator output, and disconnect the
function generator and oscilloscope.
Electrical
Operating mode, Burst, on page 1-4.
Characteristic
Checked
Equipment
One 50 W coaxial cable
Required
One oscilloscope
Prerequisites
The AWG2021 meets the prerequisites listed on page 4-7.
1. Install test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG2021 CH1 output connector
through the coaxial cable to the CH1 vertical input connector on the
oscilloscope (see Figure 4–5).
AWG2021
AWG2021

Figure 4-5: Burst Mode Initial Test Hookup

b. Set the oscilloscope controls:
Vertical
. . . . . . . . . . . . . . . . . . . . . .
CH1 coupling:
CH1 scale
. . . . . . . . . . . . . . . . .
CH1 input impedance:
Horizontal
Sweep
. . . . . . . . . . . . . . . . . . . .
Trigger
Source
. . . . . . . . . . . . . . . . . . .
Coupling
. . . . . . . . . . . . . . . . . .
Slope
. . . . . . . . . . . . . . . . . . . .
Level
. . . . . . . . . . . . . . . . . . . .
Mode
. . . . . . . . . . . . . . . . . . . .
Oscilloscope
TDS
540
CH1
. . . . . . . . . . . . .
DC
0.2 V/div.
50 W
. . . . . . .
500 ms/div.
CH1
DC
Positive
+100 mV
Norm