Instructions for use Spare cartridge for EXPERTtorque mini - 1.007.9458, 1.007.9457
5 Function test
5 Function test
▶ Insert a tool (cutter or grinder) and turn it radially by hand in different direc‐
tions, applying slight thumb pressure, to precisely align the O-rings.
▶ Start the medical device.
▶ Perform a test run under axial and radial load.
▶ Tighten the cover again applying a torque of 100 Ncm ± 6 Ncm.
Note
If the unit runs poorly or heats up in untypical manner, please return the entire
turbine contra-angle handpiece to KaVo for testing.
7/8