Troubleshooting

Problem
Will not aspirate
Piston and dispensing
cartridge difficult to move
Liquid behind the piston of
the dispensing cartridge
Aspirated liquid can't be
delivered into the bottle
Possible cause
– Dispensing cartridge not locked
– Dispensing cartridge leaks
– Filling valve leaks or is sticking
– Connection between filling tube and
tube of the cell-culture
leaks
– Tube of cell-culture
blocked or sticking
– Crystal deposition
– Improper liquid aspirated
– Dispensing cartridge leaks
– Discharge valve sticking or damaged
- 18 -
Corrective action
– Lock dispensing cartridge (⇒ page 8).
– Mount new dispensing cartridge (⇒ page 13).
– Clean filling valve (⇒ page 16) or replace valve and seal.
– Stick adapter firmly onto the filling tube and pull tube firmly
onto the adapter.
– Clean tube or replace.
– Replace dispensing cartridge (⇒ page 13).
– Check, if liquid is proper to aspirate.
– Replace dispensing cartridge (⇒ page 13).
– Clean valve (⇒ page 16) or replace.