Functional Test Of The Ek80 With Ec150-3C - Simrad EK80 Harbour Acceptance Test

Wide band scientific echo sounder
Hide thumbs Also See for EK80:
Table of Contents

Advertisement

Simrad EK80 Harbour Acceptance Test

Functional test of the EK80 with EC150-3C

A brief functional test is used to verify that the EK80 is operational with the EC150-3C.
The EC150-3C is a dual purpose unit. It can be used either as an acoustic Doppler
current profiler (ADCP) instrument to measure water current or as a split-beam echo
sounder. It can not operate these two functions simultaneously.
Prerequisites
The EK80 is installed as specified in the EK80 and EC150-3C installation manuals.
• The EK80 is turned off.
• The EC150-3C is installed in the EK80user interface.
• All relevant external sensors are connected to the EK80. The sensors are turned
on and operate normally.
• The vessel is berthed.
Neither tools nor instruments are required.
Caution
You must never set the EK80 to "ping" unless the transducer is submerged in water.
The transducer may be damaged if it transmits in open air.
Procedure
1
Make sure that the EK80 units have been set up to operate with the supply voltage
you have available.
2
Make sure that the transducer cable is not installed close to power cables.
3
Turn all EK80 system units on.
4
On the Processor Unit desktop, double-click the EK80 icon to start the program.
5
Open the
About
6
Make sure that the latest software version is installed.
7
Select operating mode: Echo sounder
a
Open the
Operation
b
Select
Normal Operation
c
To activate echo sounder operation, click the small option button on the left
side of the
d
Select
OK
8
Start normal operation.
9
Make sure that the EK80 user interface is fully operational.
30
dialog box.
menu.
.
table.
ES
to save the selected setting and close the dialog box.
401925/C