Simrad EK80 Harbour Acceptance Test page 29

Wide band scientific echo sounder
Hide thumbs Also See for EK80:
Table of Contents

Advertisement

11 Select
on the top bar.
Help
a
Make sure that the online help opens on its start page.
b
Close the online help.
12 On the
File Setup
13 Make sure that the bottom detection parameters have been set correctly.
14 Make sure that the EK80 is fully operational
15 For each channel:
a
Make sure that you can detect the bottom.
b
Start raw data recording.
c
Allow the data recording to run approximately five minutes.
d
Stop raw data recording.
e
Use a file manager, and verify that the recorded file(s) have been saved on
the chosen disk.
f
Verify that the playback is operational.
Result
Requirements
The correct AC or DC supply voltage is available.
The transducer cables are installed with a safe distance to power cables.
The latest software version is installed.
The EK80 starts up in a controlled manner.
The basic functionality of the EK80 is present.
No error messages are presented.
The context sensitive online help is functional.
The bottom is successfully detected.
Raw data can be recorded and played back.
Date and signature:
Related topics
Customer acceptance form, page 46
Secondary procedures, page 47
401925/C
page, define the recording parameters.
Testing the EK80 operational functionality
Results
29