First Commissioning; Operation; Switching On The Instrument - Dräger X-am 5000 Instructions For Use Manual

Hide thumbs Also See for X-am 5000:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 11
What is what
15
1
2
2
3
X-am 5000
7
6
4
0
5
1
Gas entry
2
Alarm LED
3
Buzzer
4
[OK] key
5
Power pack
6
[ + ] key
7
Display
Special symbols:
Fault message
Warning
Display peak value
Show TWA
Show STEL
Bump test mode
Fresh air calibration
12
ch
4
8
%LEL
O
2
Vol%
9
CO
ppm
H
S
2
ppm
10
NH
3
ppm
12
13
14
11
8
IR interface
9
Fastening clip
10
Type plate
11
Charging contacts
12
Measured gas display
13
Measured value display
14
Special symbols
15
Tool for replacing the sensor
1-button calibration
Single gas calibration
Password required
Battery 100 % full
Battery 2/3 full
Battery 1/3 full
Battery empty
Configuration
For the individual configuration of an instrument with standard configuration, the
instrument must be connected to a PC via the USB infrared cable (order no.
83 17 409) or the E-Cal system. The PC software "Dräger CC-Vision" is used to
perform the configuration.
– Changing the configuration: see Technical Manual.
Standard instrument configuration:
Dräger X-am 5000
1)
Bump test mode
1)
Fresh air calibr.
1)
Life sign
1
Switch off
1)
LEL factor
(ch
)
(4.4 vol. % corresponds to 100 %LEL)
4
1)
Averaging time
1) Different settings can be selected to meet customer requirements on delivery. The current setting
can be checked and changed with the Dräger CC Vision software.

First commissioning

Before using the instrument for the first time, insert the batteries supplied or
a charged T4 NiMH power pack T4 (HBT 0000 type, order no. 83 18 704) /
T4 HC (HBT 0100 type, order no. 83 22 244) (refer to chapter "Replacing the
batteries" on Page 15). The Dräger X-am 5000 is ready for operation.

Operation

Switching on the instrument

 Press and hold the [OK] key for approx. 3 seconds until the countdown
» 3 . 2 . 1 « shown on the display has elapsed.
– All the display segments, including the visual, audible and vibration alarms,
are activated for a short time.
– The software version is displayed.
– The instrument performs a self test.
– The next sensor to be calibrated is displayed, along with the remaining days
to the next calibration, e. g. » Ex %LEL CAL 20 «.
– The time until the bump test interval elapses is displayed in days, e.g. » bt 123 «.
Quick bump test
On
On
allowed
4.4 (vol. %)
15 minutes for STEL
8 hours for TWA