Installation
February 2020
Shorten the coaxial probe
Prerequisites
Figure 3-11: Maximum Shortening of Coaxial Probe
A. Ordered probe length (L) > 49 in. (1250 mm)
B. Ordered probe length (L) ≤ 49 in. (1250 mm)
C. Maximum shortening length: 23.6 in. (600 mm)
D. Minimum probe length: 15.7 in. (400 mm)
Procedure
1. Mark where to cut the probe.
38
•
Pipes longer than 49 in. (1250 mm) can be shortened by as much as 23.6 in.
(600 mm).
•
Pipes shorter than 49 in. (1250 mm) can be cut as long as the remaining length
is not less than 15.7 in. (400 mm).
Reference Manual
00809-0100-4308
Rosemount 3308 Series