February 2020
4. Verify the spring-loaded pin at the center of the process seal is properly inserted.
5. If the transmitter head is not mounted directly, attach the protection plug to the
6. Rotate the new transmitter head so the device display faces the desired direction.
7. Tighten the nut. Max torque is 30 ft-lb (40 Nm).
Postrequisites
Configure the transmitter.
Related information
Configuration
138
When inserted properly, only the plunger is seen above the edge inside the seal
hole.
Note
Do not remove the process seal from the adapter.
process seal to protect the exposed parts from dust and water. If a protection plug
is not available, then cover the process seal with a plastic bag.