The datum clamp fixes the RTLC40 scale rigidly to the substrate at the location chosen. The metrology of the system may be compromised if the datum clamp is not used.
The datum clamp does not need to be fitted adjacent to a reference mark. It can be positioned anywhere along the axis depending upon the customer's requirements.
Mechanical clamp
Reference mark
selector magnet
15 ±1
IN-TRAC reference mark
Bolted datum clamp
12
M3 × 6 low head screw,
supplied with clamp
†
Additional screws available (Pack of 25 A-9584-2047)
Adhesive clamp
Using dispensing tip apply Loctite
435
between scale and FASTRACK so it wicks underneath adjacent to user selected datum location as shown.
®
™
NOTE: Drawing shows the scale datum adjacent to the chosen reference mark.
QUANTiC RTLC40 / FASTRACK installation guide
2.3
1.7
5 min
thread
depth
Do not place clamp tip
over shaded area
NOTE: When reference mark and datum clamp aligned as
shown, reference output will be positionally repeatable with