Troubleshooting - Renishaw QUANTiC RTLC40 Installation Manual

Incremental linear encoder system
Hide thumbs Also See for QUANTiC RTLC40:
Table of Contents

Advertisement

Troubleshooting

Fault
LED on the readhead is Blank
LED on the readhead is Red flashing
during installation mode
Unable to get a Green LED over the
complete axis length
Can't initiate the calibration routine
During calibration the LED on the
readhead remains single flashing
Blue even after moving it along
the full axis length
During calibration the LED on
the readhead is double flashing
Blue even after moving it past the
reference mark several times
QUANTiC RTLC40 / FASTRACK installation guide
Cause
There is no power to the readhead
The signal strength is < 50%
System run-out is not within specification
Signal size is < 70%
The system has failed to calibrate the
incremental signals due to the signal
strength being < 70%
Incorrect termination (analogue variant only)
The readhead is not seeing a
reference mark
Possible solutions
u
Ensure 5 V supplied at the readhead
u
For cable variants check correct wiring of connector
u
If using the analogue variant of QUANTiC with the ADTi ensure appropriate adaptor cables are connected
u
Check the readhead optical window and scale are clean and free from contamination
u
Restore factory defaults
(page 14)
and check alignment of the readhead. In particular:
Rideheight
Yaw
Offset
u
Ensure the correct scale and readhead combination
u
Use a DTi gauge and check the run-out is within specifications
u
Restore factory defaults
(page 14)
u
Realign readhead to obtain a Green flashing LED at the mid-point of the run-out
u
Recalibrate the system
(page 13)
u
Realign readhead to obtain a green flashing LED
u
Exit CAL mode and restore factory defaults
u
(page 12)
Check readhead set-up and alignment
u
Check output signal termination
(page 19)
u
When using with ADTi-100 in stand- alone mode ensure Termination tool is connected,
Renishaw part number A-6195-2132
u
Exit CAL mode and restore factory defaults
u
Check readhead set-up and alignment
(page 12)
u
Ensure correct position of reference mark selector magnet
u
Ensure you are moving the readhead past your chosen reference mark several times
u
Check the readhead/selector magnet orientation
u
Check the readhead optical window and scale are clean and free from contamination
(page 14)
(page 14)