End Covers; Datum Clamp - Renishaw QUANTiC RTLC40-S Installation Manual

Incremental linear encoder system
Hide thumbs Also See for QUANTiC RTLC40-S:
Table of Contents

Advertisement

End covers

The end cover kit is designed to be used with RTLC40-S scale to provide protection for exposed scale ends.
NOTE: End covers are optional and can be fitted before or after readhead installation.
1. Remove the backing tape from the adhesive tape on the back of the end cover.
2. Align marker on the edges of the end cover with the end of the scale and place end cover over the scale.
*
NOTE: There will be a gap
between the end of the scale and the adhesive tape on the end cover.
*  gap

Datum clamp

IMPORTANT: Allow 24 hours after scale application before fitting datum clamp.
The datum clamp fixes the RTLC40-S scale rigidly to the substrate at the location chosen.
CAUTION: The metrology of the system may be compromised if the datum clamp is not used.
The datum clamp does not need to be fitted adjacent to a reference mark. It can be positioned
anywhere along the axis depending upon the customers' requirements.
1. Remove the backing paper from the datum clamp.
Place the datum clamp with cut-out against the scale
at the chosen location.
QUANTiC RTLC40-S installation guide
2. Place a small amount of adhesive (Loctite
in the cut-out on the datum clamp, ensuring none
of the adhesive wicks onto the scale surface.
Dispensing tips are available.
®
435
)
Ensure the adhesive wicks along
the entire length of cut-out.