IKA MATRIX DELTA PLUS Manual page 13

Table of Contents

Advertisement

/// Menu structure
Tempering
Control rate
Timer
Timer / Counter
Start Condition
A
B
Mode
A
C
...
B
C
Graph
Axis scaling
Menu
Program
Program 1
Program 2
...
Program 5
Safety
Password
Settings
Languages
Units
Display
Sound
Factory Settings
Information
24
Factory setting
Fast
Moderate
-
Slow
-
00:00:00 [hh:mm:ss]
Immediately
Set Temperature Reached
-
-
-
X-Axis
5 minutes
Y-Axis
Automatic
Manual
Maximum
110.0 °C
Minimum
-10 °C
-
-
-
-
000
English
Deutsch
-
...
-
Temperature
°C
°F
-
Background
Black
White
-
Firmware Update Information
Key Tone
Software
Version
1.02.003/1.01.021
Firmware Update Information
yes
/// Menu details
Tempering:
Control rate:
In the "Control rate" menu option, you can select the desired control rate between "Fast", "Mod-
erate" and "Slow"� A check mark (√) indicates the control rate is activated�
Fast: Fast tempering with a certain level overshooting�
Moderate: Moderate tempering with minimal overshooting�
Slow: Precise tempering without overshooting�
Timer:
In the "Timer" menu option, you can set "Timer / Counter" for "Tempering"�
You can set "Timer" value from 00:00:01 [hh:mm:ss] to 99:59:59 [hh:mm:ss]�
The tempering function stops automatically after the set time has elapsed and the set time
appears on the display�
Note: If the tempering function is stopped before the set timer value expires, the reverse
counting of the timer is interrupted�
If you set the "Timer" value to 00:00:00 [hh:mm:ss], the counter is active�
The counter automatically starts to run fom 00:00:00 [hh:mm:ss] upto 99:59:59 [hh:mm:ss]�
When the maximum value is reached, the device stops the tempering process�
In the menu option, you can set the timer / counter "Start contidion" as:
Immediately: The timer / counter starts immediately when starting the tempering function�
Set temperature reached: The timer / counter starts only when the set temperature reached�
A
B
Mode:
C
...
Mode A:
After power on / power failure no automatic restart of functions�
Mode B:
After power on / power failure automatic restart of functions, depending on previous settings�
Mode C:
Set values (set in mode A or B) cannot be changed� After power on / power failure automatic
restart of functions, depending on previous settings�
25