Measurement Application Notes

Low Loss DUT Measurements

Low loss trace ripple is a product of switch repeatability and calibration technique. Low
insertion loss measurements exhibiting a 0.1 dB uncertainty ripple residing on the measurement
response is not uncommon.

Switch Configuration Command Speed

When writing address and data values to the test set directly, include a 15 ms wait before
issuing another address and data set. The delay of 15 ms is required for the test set internal
switches to settle. Failure to add the delay could cause the test set internal switches to stick or
cause intermittent failures.

Available Options

Test Set Options
The Test Set has one available Test Set option: Refer to
• Standard - mechanical switches for low loss.
• Option 129 - adds an external source path for Ports 1, 3, 5, 7, 9, and 11.
Mounting Kits
The Test Set available cable options:
• U3022BH10-1CN Front Handle Kit 1CN105A* (5063-9229)
• U3022BH10-1CM Rack Mount Kit without Handles 1CM130A* (5063-9216)
• U3022BH10-1CP Rack Mount Kit with Handles 1CP126A* (5063-9236)

Network Analyzer Requirements

All network analyzers require Option 551 to be installed for multiport operation (N-Port error
correction and measurement capability).
analyzer option requirements. In order for the measurement system to cover the full frequency
range of the Test Set option listed above, the network analyzer's frequency range should equal
or exceed that of the Test Set.
Table 2
PNA-X Configuration Requirements
14-Port Network Analyzer System
N5241/2A/B PNA-X
24-Port Network Analyzer System
N5241/2A/B PNA-X
The Test Set files indicated in
analyzer file directory location: c:\Program Files (x86)\Agilent\Network
Analyzer\testsets
12
*
New dark color version (Keysight color palette 2015)
Table 2
(below) provides a listing of other network
Options
Test Set File
U3022BH10_pnax_p4.tsx
400
400
U3022BH10_x2_pnax_p4.tsx
Table 2
above must be installed into this network
Available Options
Figure 82 on page
104.
System Figure
1, 2