Troubleshooting The Test Set; Non-Rf Malfunctions - Keysight Technologies U3022BH10 User's And Service Manual

Multiport test set
Table of Contents

Advertisement

U3022BH10

Troubleshooting the Test Set

If the U3022BH10 is not operating properly, use the following procedures to isolate and repair
the type of failure encountered. It is recommended that a qualified service technician perform
the following procedures.
Refer to the Keysight PNA Series: Service & Support Home Page at:
http://na.support.keysight.com/pna
To request service, please contact your local service center. In the US, call 800-829-4444. For
a listing of service centers worldwide, please visit us at http://www.keysight.com/find/assist.
When sending the Test Set in for repair, please also send your PNA and cables. This will allow
us to verify proper operation of the complete system, and speed your repair
turn-around time. Refer to

Non-RF Malfunctions

If the Test Set is not operating properly, use the following checklist to aid in isolating the
source of the non-RF malfunction. Typical malfunctions may be the loss of the rear panel fan,
control lines, or front panel indicator lights.
The opening of covers or removal of parts is likely to expose the user to
dangerous voltages. Disconnect the instrument from all voltage sources
before opening.
1. Verify the front panel power switch is operational.
a. The rear panel FAN and front panel Active LED should operate when the Standby Switch is
in the ON position.
2. AC Line voltage checks (remove AC power from the instrument).
a. Ensure the proper AC line voltage is present at the instrument line cord.
b. Remove the AC power cord from the instrument. Check the instrument AC line module
fuse; refer to
Figure 7, page
c. Check the internal AC lines fuses. Remove the instrument bottom cover. Near the rear
panel are two fuse holders. Remove the the fuses and verify that they are operational. See
Figure 79, page
3. Internal DC Power Checks. During this check you will reapply AC power to the instrument with
the top cover removed. Connect the AC power cord to the Test Set.
a. Set the front panel switch to the Standby position. No fans or indicator lights should be
operational.
b. Set the front panel switch to the ON position. The rear panel and internal power supply fan
should be operational. The front panel Active LED should be ON if the Test Set is
connected to an active PNA-X.
c. Verify that the DC power indicator LEDs are ON. Refer to
Note: The LED indicators only indicate the presence of a DC voltage, not the DC voltage
level. Use the indicated wire locations for voltage measurements. TP1 is your common.
U3022BH10 User's and Service Guide
for further information.
"Keysight Support, Services, and Assistance" on page
21.
100.
Troubleshooting the Test Set
123.
Figure 87, page
110.
109

Advertisement

Table of Contents
loading

Table of Contents