Source Signal Path Insertion Loss Test - Keysight Technologies U3022BH10 User's And Service Manual

Multiport test set
Table of Contents

Advertisement

U3022BH10

Source Signal Path Insertion Loss Test

Connect the RF Flex cables to the Front Panel Test port and Source IN access port indicated in
Table 26
below. Check that the results are similar to
Table 26
Source Signal Path Test Instructions
Path
RF Path Description
#
Odd Ports
1
Source OUT to Port 3
Source OUT to Port 5
2
Source OUT to Port 7
3
Source OUT to Port 9
4
Source OUT to Port 11
5
Even Ports
6
Source OUT to Port 4
Source OUT to Port 6
7
Source OUT to Port 8
8
Source OUT to Port 10
9
Source OUT to Port 12
10
1
Use the Source IN port associated with this group of test ports.
Figure 88
Source OUT to Ports 3-12 Path Response
U3022BH10 User's and Service Guide
Figure 88
Control Mode Entry
<address>.<data>
1
0.1
0.2
0.3
0.4
0.5
16.1
16.2
16.3
16.4
16.5
Troubleshooting the Test Set
below.
Insertion Loss
Path Components
(Typical)
P3 CPLR, S10
P5 CPLR, S10
Figure 88
P7 CPLR, S10
P9 CPLR, S10
P11 CPLR, S10
P4 CPLR, S20
P6 CPLR, S20
P8 CPLR, S20
Figure 88
P10 CPLR, S20
P12 CPLR, S20
113