3. Swivel lens for magnification
There is a small magnification lens which can be swivelled by
360° on the operation otoscope with approx. 2.5-fold magnifica-
tion.
13. Replacing the lamp
L1 otoscope
Remove the specula fitting from the otoscope. Screw out the
lamp counter clockwise.
Screw in the new lamp clockwise and replace the specula fit-
ting. L2, L3 otoscopes, ri-derma, bent-arm illuminator, nasal
speculum and blade holder
Screw the instrument head off the battery holder. The lamp is
located at the base of the instrument head. Pull the lamp out of
the instrument head with thumb and forefinger or a suitable tool.
Insert a new lamp.
Ophthalmoscopes
Remove the instrument head from the battery holder.
The lamp is located at the base of the instrument head.
Remove the lamp from the instrument head with thumb and
forefinger or a suitable tool. Insert a new lamp.
Caution: The pin on the lamp must be inserted into the guide
groove on the instrument head.
Veterinary/human operation otoscope
Screw the lamp out of the fixture in the operation otoscope and
screw in a new lamp.
Instrument heads: Retinoscope slit and spot
Remove the instrument head from the battery handle. The lamp
is located in a sleeve at the base of the instrument head. Remove
the lamp from the sleeve using the thumb and index finger or
a suitable tool. Insert the new lamp firmly into the sleeve and
replace the sleeve back into the instrument head so that the base
of the lamp fits into the slot on the instrument head.
Information on care, cleaning and disinfection
All Riester ri-scope
®
instrument heads can be cleaned on the
outside with a moist cloth. Furthermore, the following disinfec-
tants can be used: Aldehydes (formaldehyde, glutaraldehyde,
aldehyde fission products) or tensides. All parts of instruments
with the exception of the swivel lens, magnifying glass and the
cover glass can also be disinfected with alcohols. When using
these substances it is absolutely essential to follow the instruc-
tions of the manufacturer. A soft and, as far as possible, lint-free
cloth or cotton bud can be used as an auxiliary aid for cleaning or
disinfection.
Attention
Never place the instrument heads in liquid. Take care that no
liquid enters inside the casing.
a) Sterilisation
According to the current school of thought (Test Centre for Me-
dical Devices in Tübingen), sterilisation is only prescribed in the
case of operative procedures.
b) Reusable ear specula
Although sterilisation is not necessary as described in a), it is
nevertheless possible. The reusable ear specula can be sterilised
at 134°C and 10 minutes retention time in a steam sterilizer.