Riester ri-former Instructions Manual page 19

Hide thumbs Also See for ri-former:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 14
Guidelines and manufacturer's declaration - electromagnetic immunity
The Diagnostic station ri-former
environment as specified below. The customer or the user of the Diagnostic station
ri-former
®
Immunity tests
Electrostatic
discharge
(ESD)
according to
IEC61000-4-2
Fast transient
electrical
interference/
bursts
according to
IEC61000-4-4
Surges
IEC61000-4-5
Voltage dips,
short-time
interruptions and
fluctuations in
the supply
voltage according
to IEC61000-4-11
Magnetic field at
the mains
frequency (50Hz)
according to
IEC61000-4-8
Note - U
is the alternating supply voltage prior to application of the test level.
T
®
is intended for operation in an electromagnetic
should ensure that it is used in such an environment.
IEC 60601 test
level
± 6 kV contact
discharge
± 8 kV air
discharge
± 2 kV for power
lines
± 1 kV for input
and
output lines
± 1 kV voltage
phase-to phase
± 2 kV voltage
phase-to earth
<5% U
T
(>95 % drop
in U
)
T
for 0.5 cycles
40% U
T
(60 % drop in U
)
T
for 5 cycles
70 % U
T
(30 % drop in U
)
T
for 25 cycles
<5% U
T
(>95 % drop
in U
)
T
for 5 s
3 A/m
Compliance
level
± 6 kV contact
discharge
± 8 kV air
discharge
± 2 kV for power
lines
± 1 kV for input
and
output lines
± 1 kV voltage
phase-to phase
± 2 kV voltage
phase-to earth
<5% U
T
(>95 % drop
in U
)
T
for 0.5 cycles
40% U
T
(60 % drop in U
)
T
for 5 cycles
70 % U
T
(30 % drop in U
)
T
for 25 cycles
<5% U
T
(>95 % drop
in U
)
T
for 5 s
3 A/m
Electromagnetic
environment
-guidelines
Floors should be
made of wood
or concrete or be
covered with
ceramic tiles. If
the floor is
covered with a
synthetic materi-
al, the relative air
humidity must be
at least 30%.
The quality of the
supply voltage
should corres-
pond to that of a
typical business
or hospital
environment.
The quality of the
supply voltage
should corres-
pond to that of a
typical business
or hospital
environment.
The quality of the
supply voltage
should corres-
pond to that of a
typical business
or hospital
environment.
If the user of the
Diagnostic sta-
®
tion ri-former
requires continu-
ed operation du-
ring power mains
interruptions, it
is recommended
that the Diag-
nostic station
ri-former
®
be
powered from an
uninterruptible
power supply or
a battery.
Power frequency
magnetic fields
should be at level
characteristic of
a typical loca-
tion in a typical
commercial or
hospital environ-
ment
19

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents