Fluke 6100B User Manual page 124

Electrical power standards
Hide thumbs Also See for 6100B:
Table of Contents

Advertisement

6100B/6105A
Users Manual
Keyword
:CURRent
:RANGe
:RANGe?
:ACTual
:TARGet
:TRIGger?
:STORe
OUTPut
[:STATe](?)
:ROSCillator
[:STATe](?)
:SENSe(?)
:DEFer(?)
[:STATe](?)
:ACTion
:RAMP(?)
:TIME(?)
:RCLock(?)
:VOLTage
:NLIMit(?)
5-20
Parameter Form
<dnpd>, <dnpd>
[<cpd> {LOW | HIGH}]
:VOLTage
<dpnd>, <dpnd>
:VOLTage(?)
[<cpd>{ LOW | HIGH }]
:UNIT?
<dnpd>,<dnpd>
:FREQuency?
<dnpd>[,<dnpd>,<dnpd>,<dnpd>,<dnpd>]
:DUMP?
<bool> {OFF | ON | 0 | 1}
<bool> {OFF | ON | 0 | 1}
<bool> {OFF | ON | 0 | 1}
<bool> {OFF | ON | 0 | 1}
[<cpd> {APPLy | UNDO}]
{FAST | SLOW}
<dnpd>
<dnpd>
<cpd> {LOW | HIGH}
Notes
Calibration range: <dnpd> = Low limit, High limit.
Calibration range query.
<dnpd> = low limit, high limit.
Response is VOLT (voltage) or CURR (current).
Query only.
<dnpd> = Amplitude, Angle. Note: Angle relative to
master phase.
Query only of Actual frequency.
<dnpd> = Target point.
Or,
<dnpd> =
Point,
Fund Freq,
Harmonic,
Amplitude,
Angle.
Note1: Angle relative to master phase.
Note2: Second form is only required when
changing target point.
Dump all stores for active range:
Point,(<target data>,<actual data>), ...
<target data> =
Fund, Harm, Ampl, Angle
<actual data> =
Freq, Ampl, Angle