Note: If one of the source parameters has been changed after using this command a valid
default will be set (typically "MAIN"); so it is recommend that the "MUT:TEST:PSOUrce"
command is sent after any "MUT:SOURce" or "REFerence:SOURce" commands, to
prevent unexpected side effects.
The source can be:
CH1 to CH6
SUM456
MEAN456
MEAN56
EMUT
MAIN

5-92. ENERgy: ABORt

This command allows an active test sequence to be aborted. Its action is similar to hitting
the 'STBY' key or sending ':OUTPut OFF', with one exception: when the 'Maintain
Voltage' output feature has been enabled (ENERgy:ENERgy:MVOLTage), the voltage
channels will remain on (even though the test sequence has been reset to the beginning
again). If ':OUTPut OFF' is used to force a test sequence to stop, all channels will turn off
regardless of whether 'Maintain Voltage' has been set."

5-93. MUT Tree

The MUT commands configure a MUT source. A MUT source is typically made up from
pulse streams from channels 1 to 3, though channels 4 to 6 can also be used when treated as
independent sources (i.e. for "drift monitoring").

5-94. MUT Meter Constant

ENERgy:MUT:CONStant(?) <dnpd>
This command specifies the meter constant (as impulses per unit energy) used by channels
considered to be MUT sources.

5-95. MUT Meter Constant Units

ENERgy:MUT:CONStant:UNIT(?){ IPWH|IPKWH|
IPMWH|WHPI|KWHPI|MWHPI }
This command specifies the meter constant units (as impulses per unit energy) used by
channels considered to be MUT sources.
IPWH
IPKWH
IPKMH
WHIP
KWHIP
KMHIP
5-58
Raw pulse streams from individual channels.
Sum of channels 4 through 5.
Arithmetic mean of channels 4 through 6.
Arithmetic mean of channels 5 and 6.
On Every MUT channel. For example, on selecting a test
duration of 100 pulses, at least 100 pulses would have to be
counted on all selected MUTs before the test sequence was
considered complete.
Equivalent pulse count of the power outputs (value calculated
from the "OUTPut:CONStant" meter constant value and active
power settings).
Impulses per Watt-hour
Impulses per killoWatt-hour
Impulses per MegaWatt-hour
Watt-hour per impulse
killoWatt-hour per impulse
MegaWatt-hour per impulse