7.8 Injection valve
Residual pulsa-
tion
Safety shutdown
Function
Maximum pres-
sure limit
Minimum pres-
sure limit
7.8
Injection valve
Actuator switch-
ing time
Maximum operat-
ing pressure
Material
7.9
Metrohm Suppressor Module (MSM)
Resistance to sol-
vents
Switching time
7.10
Metrohm CO
Material
Resistance to sol-
vents
Vacuum
Working range
Time to estab-
lish after start-
ing
Capillary volume
■■■■■■■■
96
< 1%
Automatic shutdown upon reaching the pressure limits
Adjustable from 0.1 - 50 MPa (1 - 500 bar)
■
The pump is automatically shut down at the first piston stroke
■
above the maximum limit value
Adjustable from 0 - 49 MPa (0 - 490 bar)
■
The shutdown mechanism is inactive at 0 MPa
■
The shutdown mechanism becomes active two minutes after sys-
■
tem start
The pump is automatically shut down after three piston strokes
■
below the minimum pressure limit
typ. 100 ms
35 MPa (350 bar)
PEEK
No restriction
typ. 100 ms
Suppressor (MCS)
2
Fluoropolymer
No restriction (except PFC)
Microprocessor-controlled/stabilized
< 30 s
400 µL
■■■■■■■■■■■■■■■■■■■■■■
930 Compact IC Flex SeS (2.930.1400)