Validated Reprocessing Procedure; General Safety Information; General Information; Preparations At The Place Of Use - B. Braun Aesculap Lektrafuse GN200 Instructions For Use/Technical Description

Hf generator
Hide thumbs Also See for Aesculap Lektrafuse GN200:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Start the sealing process (HF discharge) by pressing the button on the
instrument or foot switch once.
After completing the sealing process, the HF discharge is automatically
stopped.
During sealing process, the message "Sealing in Progress" is shown on
the display 5.
Fig. 12
To activate/deactivate device
Switch on/switch off the device with the power OFF/ON switch 12.
Operating mode
The Lektrafuse HF generator GN200 automatically recognizes the con-
nected instrument. The internal device settings are loaded. The user can
choose between two operating modes.
Activating the HF current
Be sure that the patient is prepared in such a way that the HF generator
can be applied without any risk to the patient.
Make sure that the required operating mode is selected.
Ensure that the instrument and foot switch, if used, are correctly con-
nected.
Activate the HF generator via the hand switch or the foot switch.
Do not activate the hand switch or foot switch continually during HF
discharge.
Note
The vessel sealing process is started and ended by pressing the foot switch
or hand switch on the instrument once. After completing the sealing pro-
cess, the device stops the HF discharge automatically. The sealing process
can be interrupted by repeatedly pressing the button.
5.

Validated reprocessing procedure

5.1

General safety information

Note
Adhere to national statutory regulations, national and international stan-
dards and directives, and local, clinical hygiene instructions for sterile pro-
cessing.
Note
For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possi-
ble variants of CJD, observe the relevant national regulations concerning
the reprocessing of products.
Note
Mechanical reprocessing should be favored over manual cleaning as it gives
better and more reliable results.
Note
Successful processing of this medical device can only be ensured if the pro-
cessing method is first validated. The operator/sterile processing technician
is responsible for this.
Note
If there is no final sterilization, then a virucidal disinfectant must be used.
Note
For up-to-date information about reprocessing and material compatibility,
see B. Braun eIFU at eifu.bbraun.com
The validated steam sterilization procedure was carried out in the Aesculap
sterile container system.
5.2

General information

Dried or affixed surgical residues can make cleaning more difficult or inef-
fective and lead to corrosion. Therefore the time interval between applica-
tion and processing should not exceed 6 h; also, neither fixating pre-
cleaning temperatures >45 °C nor fixating disinfecting agents (active
ingredient: aldehydes/alcohols) should be used.
Excessive measures of neutralizing agents or basic cleaners may result in
a chemical attack and/or to fading and the laser marking becoming
unreadable visually or by machine for stainless steel.
Residues containing chlorine or chlorides e.g. in surgical residues, medi-
cines, saline solutions and in the service water used for cleaning, disinfec-
tion and sterilization will cause corrosion damage (pitting, stress corro-
sion) and result in the destruction of stainless steel products. These must
be removed by rinsing thoroughly with demineralized water and then dry-
ing.
Additional drying, if necessary.
Only process chemicals that have been tested and approved (e.g. VAH or
FDA approval or CE mark) and which are compatible with the product's
materials according to the chemical manufacturers' recommendations
may be used for processing the product. All the chemical manufacturer's
application specifications must be strictly observed. Failure to do so can
result in the following problems:
Optical changes of materials, e.g. fading or discoloration of titanium or
aluminum. For aluminum, the application/process solution only needs
to be of pH >8 to cause visible surface changes.
Material damage such as corrosion, cracks, fracturing, premature aging
or swelling.
Do not use metal cleaning brushes or other abrasives that would dam-
age the product surfaces and could cause corrosion.
Further detailed advice on hygienically safe and material-/value-pre-
serving reprocessing can be found at www.a-k-i.org, link to "AKI-
Brochures", "Red brochure".
5.3

Preparations at the place of use

Remove any visible surgical residues to the extent possible with a
damp, lint-free cloth.
Transport the dry product in a sealed waste container for cleaning and
disinfection within 6 hours.
5.4

Preparation before cleaning

Disconnect the unit from mains power.
Remove accessories.
en
9

Advertisement

Table of Contents
loading

Table of Contents