Note
For DC power:
Power requirements assume a single transmitter per cable.
At startup, the power source must provide a minimum of 1.5 amps of short-term current per transmitter
and not pull voltage below 18 VDC.
Length and conductor diameter of the power cable must be sized to provide 18 VDC minimum at the
power terminals, at a load current of 0.7 amps.
Cable sizing formula
M = 18V + (R x L x 0.5A)
M: minimum supply voltage
R: cable resistance (in Ω/ft)
L: cable length (in ft)
Typical power cable resistance at 68 °F (20.0 °C)
Wire gauge
14 AWG
16 AWG
18 AWG
20 AWG
2
2.5 mm
2
1.5 mm
2
1.0 mm
2
0.75 mm
2
0.50 mm
2.3.1 Maximum cable lengths between sensor and transmitter
The maximum cable length between the sensor and transmitter that are separately installed is determined by
cable type.
Cable type
Micro Motion 4-wire remote mount
Micro Motion 9-wire remote mount
User-supplied 4-wire
Installation Manual
Resistance
0.0050 Ω/ft
0.0080 Ω/ft
0.0128 Ω/ft
0.0204 Ω/ft
0.0136 Ω/m
0.0228 Ω/m
0.0340 Ω/m
0.0460 Ω/m
0.0680 Ω/m
Wire gauge
Not applicable
Not applicable
VDC 22 AWG (0.326 mm²)
August 2020
Maximum length
1,000 ft (305 m) without Ex-
approval
500 ft (152 m) with IIC rated
sensors
1,000 ft (305 m) with IIB rated
sensors
60 ft (18 m)
300 ft (91 m)