Load Match, Insertion Loss, Isolation For Switch 9 (Spdt, Port 9, 9A And 9B); Description; Specification - Agilent Technologies ENA Series Service Manual

Configurable multiport test set
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Performance Tests
11. LOAD MATCH, INSERTION LOSS, ISOLATION for SWITCH 9 (SPDT, PORT
9, 9A and 9B)
11. LOAD MATCH, INSERTION LOSS, ISOLATION for
SWITCH 9 (SPDT, PORT 9, 9A and 9B)

Description

This test checks the Load Match, Insertion Loss and Isolation for ports 9COM, 9A and 9B
of the E5092A. These parameters checked at 50 MHz through 20 GHz with the N5230A
Network Analyzer.

Specification

@ 23°C ± 5°C (Warming up over 90 minutes)
Load Match (Selected port)
SPDT Switch
50 MHz to 2 GHz
2 GHz to 4 GHz
4 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 18 GHz
18 GHz to 20 GHz
Load Match (Unselected port)
SPDT Switch
50 MHz to 3 GHz
3 GHz to 10 GHz
10 GHz to 16 GHz
16 GHz to 18 GHz
18 GHz to 20 GHz
Load Match (Common port)
SPDT Switch
50 MHz to 2 GHz
2 GHz to 4 GHz
4 GHz to 8 GHz
8 GHz to 10 GHz
38
Specification
17 dB
11 dB
8 dB
7 dB
4 dB
4 dB
Specification
17 dB
11 dB
8 dB
6 dB
4 dB
Specification
16 dB
11 dB
8 dB
7 dB
Chapter 2