Performance Tests
3. LOAD MATCH, INSERTION LOSS, ISOLATION TEST for SWITCH 1 (SP4T,
Port 1, 1A, 1B, 1C and 1D)
3. LOAD MATCH, INSERTION LOSS, ISOLATION TEST
for SWITCH 1 (SP4T, Port 1, 1A, 1B, 1C and 1D)

Description

This test checks the Load Match, insertion Loss and Isolation for ports 1A, 1B, 1C and 1D
of the E5092A. These parameters checked at 50 MHz through 20 GHz with the N5230A
Network Analyzer

Specification

@ 23°C ± 5°C (Warming up over 90 minutes)
Load Match (Selected port)
SP4T Switch
50 MHz to 2 GHz
2 GHz to 4 GHz
4 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 18 GHz
18 GHz to 20 GHz
Load Match (Unselected port)
SP4T Switch
50 MHz to 3 GHz
3 GHz to 10 GHz
10 GHz to 16 GHz
16 GHz to 18 GHz
18 GHz to 20 GHz
Load Match (Common port)
SP4T Switch
50 MHz to 2 GHz
2 GHz to 4 GHz
4 GHz to 8 GHz
8 GHz to 10 GHz
14
Specification
17 dB
11 dB
8 dB
7 dB
4 dB
4 dB
Specification
17 dB
11 dB
8 dB
6 dB
4 dB
Specification
16 dB
11 dB
8 dB
7 dB