Load Match (COM)
Frequency
Port
Range [Hz]
(Selected
Port)
7COM
50M to 2G
(7A)
2G to 4G
4G to 8G
8G to 10G
10G to 20G
50M to 2G
7COM
(7B)
2G to 4G
4G to 8G
8G to 10G
10G to 20G
Insertion Loss
From Port
Frequency
- To Port
Range [Hz]
7COM - 7A
50M to 100M
100M to 2G
2G to 3G
3G to 4G
4G to 6G
6G to 8G
8G to 10G
10G to 14G
14G to 18G
18G to 20G
Chapter 2
E5092A Performance Test Record
Test Limit
Test Result [dB]
[dB]
< -16.0
< -11.0
< -8.0
< -7.0
< -4.0
< -16.0
< -11.0
< -8.0
< -7.0
< -4.0
Test Limit
Test Result [dB]
[dB]
> -4.0
> -3.5
> -4.5
> -5.0
> -5.5
> -7.0
> -8.0
> -8.5
> -10.0
> -11.5
Performance Tests
Measurement
Uncertainty
[dB]
± 0.30
± 0.23
± 0.12
± 0.13
± 0.10
± 0.30
± 0.23
± 0.12
± 0.13
± 0.10
Measurement
Uncertainty
[dB]
± 0.27
± 0.24
± 0.18
± 0.19
± 0.11
± 0.09
± 0.24
± 0.23
± 0.29
± 0.59
119