Replacement Procedure
Replacement Procedure for E5092A-020

Switch Indicator LED PCA Removal

Tools Required
Procedure
Refer to Figure 5-12 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 178.
Step 2. Detach the front chassis as described in "Front Chassis Detachment" on page 179.
Step 3. Remove the front frame as described in "Front Frame Removal" on page 180.
Step 4. Remove the main board as described in "Main Board Removal" on page 184.
Step 5. Disconnect the flat cable assembly (item 1) connected to the switch indicator LED PCA.
Step 6. Remove the two TORX T10 screws (item 2 and item 3) securing the switch indicator LED
PCA to the front panel..
Step 7. Remove the switch indicator LED PCA from the assembly (item 4).
Figure 5-12
Switch Indicator LED PCA Removal
192
Box wrench 5.5mm
TORX screwdriver, T10, T15 and T20.