Performance Tests
E5092A Performance Test Record
Load Match, Insertion Loss and Isolation Test for Switch 8 (SPDT,
8COM, 8A and 8B)
Load Match (Unselected)
Port
Frequency
(Selected
Range [Hz]
Port)
8A (8B)
50M to 3G
3G to 10G
10G to 16G
16G to 18G
18G to 20G
Load Match (Selected)
Frequency
Port
Range [Hz]
(Selected
Port)
8A (8A)
50M to 2G
2G to 4G
4G to 8G
8G to 10G
10G to 18G
18G to 20G
8B (8B)
50M to 2G
2G to 4G
4G to 8G
8G to 10G
10G to 18G
18G to 20G
122
Test Limit
Test Result [dB]
[dB]
< -17.0
< -11.0
< -8.0
< -6.0
< -4.0
Test Limit
Test Result [dB]
[dB]
< -17.0
< -11.0
< -8.0
< -7.0
< -4.0
< -4.0
< -17.0
< -11.0
< -8.0
< -7.0
< -4.0
< -4.0
Measurement
Uncertainty
[dB]
± 0.43
± 0.37
± 0.28
± 0.28
± 0.35
Measurement
Uncertainty
[dB]
± 0.37
± 0.27
± 0.18
± 0.28
± 0.29
± 0.29
± 0.37
± 0.27
± 0.18
± 0.28
± 0.29
± 0.29