Control Line Maximum Current Test - Agilent Technologies ENA Series Service Manual

Configurable multiport test set
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

From Selected Port To Unselected Port
Frequency
From Port
Range [Hz]
- To Port
(Selected
Port)
10A - 10B
50M to 500M
(10A)
500M to 1G
1G to 2G
2G to 6G
6G to 10G
10G to 18G
18G to 20G
10B - 10A
50M to 500M
(10B)
500M to 1G
1G to 2G
2G to 6G
6G to 10G
10G to 18G
18G to 20G

Control Line Maximum Current Test

Control Line
Lower Test
Group
Limit [mA]
≥ 50
A
≥ 50
B
≥ 0.5
C
≥ 0.5
D
Chapter 2
E5092A Performance Test Record
Test Limit
Test Result [dB]
[dB]
< -65.0
< -80.0
< -85.0
< -90.0
< -85.0
< -75.0
< -65.0
< -65.0
< -80.0
< -85.0
< -90.0
< -85.0
< -75.0
< -65.0
Test Result [mA]
Performance Tests
Measurement
Uncertainty
[dB]
± 0.35
± 0.49
± 0.62
± 0.71
± 0.71
± 0.44
± 0.36
± 0.35
± 0.49
± 0.62
± 0.71
± 0.71
± 0.44
± 0.36
Measurement
Uncertainty [mA]
± 0.008856
± 0.008856
± 0.000046
± 0.000046
133