Technical Specifi Cations - Atmos A 161 Operating Instructions Manual

Aspirator
Hide thumbs Also See for A 161:
Table of Contents

Advertisement

12.0
Technical specifi cations
Air fl ow rate of pump
Max. vacuum
Vacuum readout
(ATMOS C 161 / C 261 Aspirator)
Additional air regulation
(ATMOS C 161 / C 261 Aspirator)
Additional air regulation
(ATMOS A161 / A 261 Aspirator)
Collection canister
Suction hose
Voltage
Current input (max.)
Power consumption
Power cable
Operating time
Fuse
Protective earth conductor resistance
Earth leakage current
Enclosure leakage current
Patient leakage current
Heat emission
Noise level
(ATMOS C161 / C 261 Aspirator)
Noise level
(ATMOS A161 / A 261 Aspirator)
Ambient conditions
Transport/storage
Operation
Dimensions HxBxT
Weight
Regular safety relevant
inspections
Protection class (EN 60601-1)
Applied Part
Degree of protection
Classifi cation acc. to Annex IXEEC directions 93/42/EEC
CE marking
Rules applied
UMDNS-Code
GMDN-Code
* 1 bar ≈ 750,06 mm Hg ≈ 1000 hPa / depends on daily atmospheric pressure
ATMOS
A / C 161 Aspirator
®
16 ± 2 l/min
-80 kPa* ± 2 kPa
-1...0 bar (± 25 mbar) (mm Hg; kPa)*
mechanical regulating valve
Three-steps-adjustment
*
low= -25 kPa
, medium= -55 kPa
* depends on daily atmospheric pressure and Ambient conditions
1l DDS secretion canister, 1.5l Receptal
1l Medi-Vac
canister system or 1l Serres
®
ø 6 mm, 1.30 m length
230 V~ (+/-10%) 50/60 Hz
ca. 0.85 A (230 V~)
ca. 170 VA (230 V~)
2 m
Interrupted use over approx. 45 min at 230 V~, 50 Hz, 20°C
cooling period approx. 60 min., depending on ambient temperature
(Interrupted use over minimum 5 min. at 10% higher rated voltage,
60 Hz, 35°C ambient temperature, 20 min. break)
T 1.25 A/H (230 V~)
N.C. < 0,1 mA
100 J/s
50.0 dB (A) @ 1m (nach ISO 7779)
53.0 dB (A) @ 1m (nach ISO 7779)
-30...+50°C
5...90 % humidity,
non-condensing air pressure 700...1060 hPa
+10...+35°C
20...80 % humidity,
non-condensing air pressure 700...1060 hPa
250 x 255 x 180 mm
3.5 kg
recommended: once a year
II
type B
IPX 0
IIa
CE 0124
EN 60601-1: 1990 + A1: 1993 + A2: 1995
EN 60601-1-2: 2001
EN ISO 10079-1: 1999
10-219 (tracheal suction device)
37783
Issue of Technical Specifi cations: 04.07.2011
ATMOS
A / C 261 Aspirator
®
26 ± 2 l/min
-78 kPa* ± 2 kPa
*
*.
, high= -75kPa
canister system,
®
canister system
®
ca. 0,46 A (230 V~)
ca. 105 VA (230 V~)
T 1,25 A/H (100-230 V~)
21