3.2.2 During ongoing program

3.2.3 After program completion

3.3 Cooling

12
WARNING!
If the red indicator lights up an error has
occurred. The process is cancelled and
must be repeated.
WARNING!
The goods may be hot at the end of the program.
If the temperature of the goods is above 60°C, the
display shows U7.
1.
When the program is complete:
The green light at
The door unlocks automatically.
Cooling/Ventilation of goods may continue after
completed program (option) and is thus terminated
when the door is opened.
2.
Open the door and remove the goods.
A ventilator starts when the disinfection is complete:
Warm air is sucked out of the chamber.
The items to be washed are cooled using the following method:
Internal cooling: items to be washed are cooled using water
from a tank. This type of cooling does not fully meet EN ISO
15883.
Fan (option): items cooled by cold air from an external fan.
Cooling/Ventilation can continue after completed process (option).
When the green lamp is lit the cooling/ventilation can be terminated
by opening the door.
lights up.