First-Time Operation - Atmos Atmoforte 350 Operating Instructions Manual

Table of Contents

Advertisement

3.

First-Time Operation

Fig. 4.
Lid system
Locking bow
Knurled screw for removal of the lid insert
Release knob
Fill-level sensor with foam protection
Lid rim
Aperture for double socket nipple
Contacts for monitoring of the fill level
Fig. 5.
Mounting the lid
This section decribes
– how to handle the lid system of the collection jar
– how to close and insert the collection jar
– how and where to connect the tubes
– how to connect the Atmoforte 350/Record 500 to
the power line
Before putting the device into operation for the
first time, do not fail to read section 1.3 "For your
Safety".
The lid system must seal the collection jar tight to allow
the vacuum to build up inside. Fig. 4 shows the lid system
with the locking bow open.
When dealing with heavily foaming secretions,
the foam protector should be attached to the fill-level
sensor.
* Slide the lid system onto the collection jar as shown in
Fig. 5, taking care that the rim of the lid ( , Fig. 4) is
placed below the rim of the jar, and press down the
locking bow until you hear it click into place.
9