104 (Extended Device Status)
Device_Status
Mass_flowrate_exceeds_
limits_0_0
Volume_flowrate_exceeds_
limits_0_1
Simulation_is_on_0_2
Flowrate_to_zero_0_3
Maintenance_interval_is_
reached_0_4
All_totalizer_stop_0_5
Totalizer_reset_0_6
Display_value_is1600h_at_
Qmax_0_7
Device_not_calibrated_1_0
Sensor_memory_
defective_1_1
NV_data_defect_Data_
storage_1_2
No_Frontend_Board_
detected_1_3
FEB_communication_
error_1_4
Incompatible_Frontend_
Board_1_5
NV_chips_defect_on_
Motherboard_1_6
Pulse_output_is_cutted_
off_1_7
Current_output_3132_is_
saturated_2_0
CurrOut_V1V2_V3V4_
saturated_2_1
CurrOut_3132_com_
error_2_2
Option_Card_1_com_
error_2_3
Option_Card_2_com_
error_2_4
Safety_Alarm_
CurrOut_3132_2_5
CurrOut_3132_not_
calibrated_2_6
CurrOut_V1V2_not_
calibrated_2_7
CurrOut_V3V4_not_
calibrated_3_0
MB_voltages_outside_
range_3_1
An_alarm_is_simulated_3_2
Communication_card_not_
responding_3_3
Reserved_0
Coil_regulation_error_3_5
Coil_wiring_Detection_3_6
Coil Impedance
measurement_3_7
Data Type
Display
Bytes
BYTE
Hex
1
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
1
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
1
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
1
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
1
Access
104 (Extended Device Status)
Read Only
Electrode short circuit
detection_4_0
Read Only
Electrode open circuit
detection_4_1
Read Only
DC Feedback Regulation
Error_4_2
Read Only
Monitoring Comm ADC and
Read Only
RX210_4_3
Read Only
Coil Isolation_4_4
Gas bubble alarm_4_5
Read Only
Conductivity exceeds
Read Only
limits_4_6
Read Only
Sensor Temperature exceeds
limits_4_7
Read Only
TFE alarm_5_0
Read Only
EPD alarm_5_1
ADC Signal overrange_5_2
Read Only
SIL self check alarm_5_3
Read Only
Inhouse temeprature exceeds
limits_5_4
Read Only
Reserved3_5_5
Reserved4_5_6
Read Only
Reserved5_5_7
Reserved6
Read Only
Table 9 Assembly/Instance 104, Extended Device Status
Read Only
110 (Output)
Read Only
DO_Function_Activation
Read Only
Empty
Empty
Read Only
Empty
Empty
Read Only
Empty
Read Only
Empty
Empty
Read Only
Empty
DO_Flow_To_Zero
Read Only
DO_System_Zero_Adjust
Read Only
DO_Counter_Reset
DO_Counter_Stop
Read Only
DO_Dual_Range_Mass
DO_Dual_Range_Volume
Read Only
DO_Batch_Start_Stop
Read Only
Reserved1
Read Only
Empty
Table 10 Assembly/Instance 110, Output
Read Only
Read Only
Read Only
Read Only
Data Type
Display
Bytes
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
BOOL
Decimal
SINT
Hex
Data Type
Display
Bytes
BOOL
Hex
BOOL
Hex
BOOL
Hex
BOOL
Hex
BOOL
Hex
BOOL
Hex
BOOL
Hex
BOOL
Hex
1
BYTE
Hex
1
BOOL
Hex
BOOL
Hex
BOOL
Hex
BOOL
Hex
BOOL
Hex
BOOL
Hex
BOOL
Hex
BOOL
Hex
1
BYTE
Hex
1
19
Access
Read Only
Read Only
Read Only
Read Only
Read Only
Read Only
Read Only
1
Read Only
Read Only
Read Only
Read Only
Read Only
Read Only
Read Only
Read Only
1
Read Only
1
Read Only
Access
Read/Write
Read/Write
Read/Write
Read/Write
Read/Write
Read/Write
Read/Write
Read/Write
Read/Write
Read/Write
Read/Write
Read/Write
Read/Write
Read/Write
Read/Write
Read/Write
Read/Write
Read/Write