Display Analysis Results - SYSMEX XE-5000 Instruction Manual

Automated hematology analyzer
Hide thumbs Also See for XE-5000:
Table of Contents

Advertisement

6.21 Display analysis results

Manual
Next No.123456789012346
C D N R
DP No.
123456789012345
Ready
POS ERR
PNo. 123456789012345
RBC
3.41 x10^6/uL
WBC & 14.08 x10^3/uL
HGB
9.9 g/dL
NEUT
10.37
HCT
31.3 %
LYMPH
2.19
MCV
91.8 fL
MONO
1.38
MCH
29.0 pg
EO
0.03
MCHC
31.6 g/dL
BASO
0.11
PLT
191 x10^3/uL
IG
0.05
RET%
38.7 %
HPC
RET# 0.1320 x10^6/uL
NRBC
1.29 x10^3/uL
QC
AutoRinse
Maint
Sysmex XE-5000 Instructions for Use
Do not touch or remove the cover of the CP Unit
during sampler analysis.
You could be injured by the mechanical parts.
(If the cover is opened, analysis will be interrupted by
the monitoring switch.)
11. When the READY LED is lit, remove the remaining rack in
the analysis line, prepare the next samples, and repeat
the steps 8-10.
If a message calling for reagent replacement is
displayed during analysis, replace with new reagent. If
the replace reagent sequence is executed when
reagent amount is insufficient, bubbles may occur,
resulting in higher background value.
The analysis results – numeral information of latest samples -
Num
can be displayed on the Main Menu screen of the Main Unit
DP
Xm
LCD.
123456-01
The Main Menu screen is made up of five pages that can be
73.6 %
15.6 %
switched using
9.8 %
0.2 %
0.8 %
0.6 %
x10^3/uL
Reagent
• The Positive/Negative display is intended for use only
in the clinical laboratory and are not for patient
diagnosis.
• POSITIVE or ERROR judgments indicate the
possibility of sample abnormality. Such results
should be reviewed carefully and may require further
examination in accordance with the protocol of your
laboratory.
Caution!
Information
and
keys.
Caution!
CHAPTER 6 Operation
6-25