Printing Of Results - SYSMEX XW-100 Instructions For Use Manual

Automated hematology analyzer
Hide thumbs Also See for XW-100:
Table of Contents

Advertisement

Type of
Description
Histogram
Flag
PLT
Relative
frequency for
LOW
discriminator
(LD) exceeds the
range
PLT
Relative
frequency for
UPPER
discriminator
(UD) exceeds the
range
PLT
Two or more
peaks in the
histogram
PLT
Particle
distribution width
error for 20%
frequency with
the peak taken as
100%. When the
20% frequency
does not cross the
histogram two
times, this flag is
appears.

Printing of Results

• It is normal for results not to be displayed. Results are printed only to avoid the
possibility of transcription errors.
• Results printouts will list patient ID and DOB, Operator ID and time of test as well as
results for each parameter and its appropriate reference range. The printout may have
flags which should be interpreted by the Clinician. Clinicians must consult the XW- 100
Clinician's Quick Guide in this manual for additional information.
• All results interpretation must be made by the Clinician.
List of PLT Error Flags
Probable Sample Cause
Effects of cryoglobulins,
fragmented RBC or cellular
fragments of WBcs, electronic
noise.
Increase of large PLT, inclusion
of fragmented RBCs, precipitation
of cryoglobulins, PLT aggregation
or agglutination, presence of
micro-erythocytes, electronic
noise.
PLT aggregation, low PLT count
Inclusion of fragmented RBCs,
non-uniformity in PLT size,
effects of cryoglobulins.
Suggested
Action
•Retest the same
sample. If the
retest confirms
the flag, send the
sample to a
reference lab.
Same as above
Same as above
Same as above