Mounting The Fu20-Fts/Mts In Pr10 Retractable - YOKOGAWA FU20-FTS Instruction Manual

Differential ph/orp sensor
Hide thumbs Also See for FU20-FTS:
Table of Contents

Advertisement

3.5 Mounting the FU20-FTS/MTS in
PR10 retractable
1 Take the sensor out of the box and
apply Teflon tape to the appropriate
threaded end.
2 Bind the separate wires of the cable
together with a piece of tape.
3 Take the fitting out of the box and
remove the option(s), if necessary.
4 Release the pigtail (cable gland)
completely. Do not undo the part in the
metal tube!
Figure 6: Dimensional drawing PR10...-D32 with mounted FU20 sensor
5 Lead the sensor cable through the tube
of the fitting, from the side where the
knurled knob has been removed. Attach
the sensor and cable as usual.
6 Hold the sensor still and turn the metal
tube onto the sensor. Don't rotate the
cell, but rotate the tube of the fitting,
because the cable can be disconnected
from the cell, when rotating it.
7 Lead the loose part of the pigtail onto
the cable and screw it onto the fixed
part.
8 Remove the tape.
IM 12B06J03-05EN-P