Yokogawa products and parts are information for the installation and use guaranteed free from defects in of the FU20-FTS and FU20-MTS, four- workmanship and material under normal in-one wide body pH sensors. Like the use and service for a period of (typically) 12...
1.4 Serial number The Serial number is defined by nine (9) alphanumeric characters: X 1 X 2 Production location X 3 X 4 Year/Month code X 5 X 6 X 7 X 8 X 9 Tracking number Example: N3P600028 Method used for year/month numbering Table 1: Production Year code Table 2: Production Month code Year...
Ui = 18 V; Ii = 170 mA; Pi = 0.4 W; Li = 0 mH; Ci = 0 nF (VP type) Certified intrinsically safe Yokogawa pH/ORP transmitter Model FLXA21 series or Model FLXA202 series. Special conditions (X): T6 for Tamb. -40 °C to +40 °C T4 and T5 for Tamb.
3. INSTALLATION OF FU20-FTS/MTS 3.1.1 Preparing the sensor for use For optimum measurement results, the Remove the sensor from its shipping box FU20-FTS/MTS should be installed in and slide of the so-called ‘wet pocket’, the a location that offers an acceptable tube filled with solution to prevent drying out representation of the process composition of the measuring elements during shipment...
3.2 Mounting FU20-FTS/MTS using 3 Apply Teflon tape to the appropriate quick removal adapters threaded end of the sensor; The FU20-FTS/MTS can also be mounted 4 Place the O-ring and screw the mating using part of the adapter on the sensor; one of the optional quick-removal adapters 5 Mount the sensor in the adapter, making (/NSS, /NTI, /BSS, /BTI, see Figure 3).
3.3. Mounting the FU20-FTS/MTS in 5 Lead the sensor cable through the tube PR10 retractable of the fitting, from the side where the 1 Take the sensor out of the box and apply knurled knob has been removed. Attach Teflon tape to the appropriate threaded the sensor and cable as usual.
Page 13
FU20-FTS/MTS + FD40 FU20-FTS/MTS + ADAPTER Fitting Process Flow /HCNF, SPRAY CLEANER O-ring Cleaning Adapter Solution ( /FPS ) (water, acid, etc.) FU20 O-ring Figure 5. Installation examples for the FU20 IM 12B06J03-05E-E...
5. WIRING 5.1 Connections of VP with cable WU-V-D and WE10-H-D VP dual coax The numbers in the table refer to the numbers of the terminals in a Yokogawa pH instrument. Differential pH + pNA + ORP(rH) 15 16 13 17 11 12 14...
Transparent Coax 2 Note: For ORP measurement the EXA/FLXA terminals 13 and 14 should be interconnected. For ORP pH compensated measurement the EXA/FLXA terminals 14 and 15 should be interconnected. 5.4 Settings for Yokogawa pH analyzer Function pH compensated pH&ORP pH&rH...
2. Rinse sensor thoroughly with tap water For the laboratory measurement, we advice (DO NOT use demineralized water); to use the Yokogawa PH72 personal pH 3. Immerse the sensor in the first buffer meter. solution (pH 7.00 is recommended);...
6.3 Calibration of ORP and rH 6.4 Maintenance of the FU20-FTS/MTS measurements sensor For calibration of ORP and rH, the A pH sensor requires routine maintenance procedure for MANUAL CALIBRATION can to keep the measuring elements clean and be used as described in the Instruction functioning.
7. MODEL CODES Model Code Suffix Option Description FU20 Wide Body sensor Cable length 3 m cable 5 m cable not available for FTD, MTS and FTS 10 m cable 20 m cable No Cable; Vario Pin connector not avaible for MTS No Cable;...
O O O O O O Note: Information in this list is based on our general experience and literature data and given in good faith. However Yokogawa is unable to accept responsobility for claims related to this information. IM 12B06J03-05E-E...
Page 24
Yokogawa has an extensive sales and YOKOGAWA ELECTRIC CORPORATION YOKOGAWA ELECTRIC ASIA Pte. LTD. World Headquarters 5 Bedok South Road distribution network. 9-32, Nakacho 2-chome, Musashino-shi Singapore 469270 Tokyo 180-8750 Singapore Please refer to the European website www.yokogawa.com/sg Japan www.yokogawa.com (www.yokogawa.com/eu) to contact your...