Change 19
Change the Measurement Uncertainty value of 4 m Impedance Measurement Accuracy Test
for 4 m DC Resistance Measurement Accuracy Test (Opt. 001.Only) (Page 9-45) to the
following information.
Standard
100 mohm
100 kohm
Copyright 2007 Agilent Technologies
C
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Test Limits
Test Result
±1.29 mohm
±0.87 kohm
Measurement
1
Uncertainty
mohm
± 0.030 mohm
kohm
± 0.013 kohm