Agilent 4263B LCR Meter Operation Manual
Manual Change
Agilent Part No. N/A
July 2007
Change 1
Add the following steps to the step 2 of DC Bias Level Accuracy Test (page 9-9).
2-a. Set the test signal frequency to 100 kHz using Freq.
2-b. Set the test signal level to 50 mV using Level.
Change 2
Change the figure 9-4 of DC Bias Level Accuracy Test Setup without the Interface Box (page 9-10) as
follows.