Agilent Technologies 4263B Operation Manual page 2

Lcr meter
Hide thumbs Also See for 4263B:
Table of Contents

Advertisement

Change 1
Change the Measurement Uncertainty value of Test Signal Frequency Accuracy Test (Page
9-38) to the following information.
Test Signal
Frequency
100 Hz
120 Hz
1 kHz
10 kHz
20 kHz
100 kHz
Change 2
Change the Measurement Uncertainty value of Test Signal Level Accuracy Test (Page 9-38)
to the following information.
Test Signal
Level
50 mV
50 mV
50 mV
250 mV
250 mV
250 mV
1 V
1 V
1 V
Copyright 2007 Agilent Technologies
C
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Agilent 4263B LCR Meter Operation Manual
Manual Change
Test Limits
±0.010 Hz
±1.200 Hz
±0.00010 kHz
±0.0010 kHz
±0.0010 kHz
±0.010 kHz
Test Signal
Test Limits
Frequency
100 Hz
±15.0 mV
2
20 kHz
±15.0 mV
100 kHz
±15.0 mV
100 Hz
±35.0 mV
20 kHz
±35.0 mV
100 kHz
±35.0 mV
100 Hz
±0.11 V
20 kHz
±0.11 V
100 kHz
±0.11 V
1
Test Result
Measurement
Uncertainty
Hz
± 0.00031 Hz
Hz
± 0.00034 Hz
kHz
± 0.0000029 kHz
kHz
± 0.000019 kHz
kHz
± 0.000020 kHz
kHz
± 0.00019 kHz
1
Test Result
Measurement
Uncertainty
Agilent Part No. N/A
July 2008
± 0.90 mV
± 0.078 mV
± 0.18 mV
± 2.50 mV
± 0.35 mV
± 1.88 mV
± 0.00349 V
± 0.00148 V
± 0.00109 V

Advertisement

Table of Contents
loading