Agilent Technologies 4263B Operation Manual page 7

Lcr meter
Hide thumbs Also See for 4263B:
Table of Contents

Advertisement

Change 14
Change the Measurement Uncertainty value of 1 m Impedance Measurement Accuracy Test
for 1 m DC Resistance Measurement Accuracy Test (Opt. 001 Only) (Page 9-43) to the
following information.
Standard
100 mohm
100 kohm
Change 14
Change the Measurement Uncertainty value of 2 m Impedance Measurement Accuracy Test
for 2 m Capacitance Measurement Accuracy Test (Page 9-44) to the following information.
Test Signal
Standard
Frequency
100 pF
100 pF
100 pF
20 kHz
1 nF
1 μF
1 μF
1 μF
Change 15
Change the Measurement Uncertainty value of 2 m Impedance Measurement Accuracy Test
for 2 m Resistance Measurement Accuracy Test (Page 9-44) to the following information.
Test Signal
Standard
Frequency
100 mohm
100 mohm
Copyright 2007 Agilent Technologies
C
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Test Limits
±1.05 mohm
±0.87 kohm
Parameter
Measured
1 kHz
Cp
D
10 kHz
Cp
D
2
Cp
D
100 Hz
Cp
D
100 Hz
Cp
D
1 kHz
Cp
D
10 kHz
Cp
D
Parameter
Measured
100 Hz
R
1 kHz
R
Measurement
1
Test Result
Uncertainty
mohm
± 0.030 mohm
kohm
± 0.014 kohm
Test Limits
Test Result
±0.19 pF
±0.0019
±0.51 pF
±0.0051
±1.10 pF
±0.0110
±0.002 nF
±0.0020
±0.0018 μF
±0.0018
±0.0011 μF
±0.0011
±0.0026 μF
±0.0026
Test Limits
Test Result
±0.68 mohm
mohm
±0.72 mohm
mohm
Measurement
1
Uncertainty
± 0.011 pF
± 0.000073
± 0.013 pF
± 0.00014
± 0.022 pF
± 0.000094
± 0.000090 nF
± 0.000075
± 0.000077 μF
± 0.000047
± 0.000069 μF
± 0.000033
± 0.00014 μF
± 0.000062
Measurement
1
Uncertainty
± 0.030 mohm
± 0.034 mohm

Advertisement

Table of Contents
loading

Table of Contents