Change 3
Change the Measurement Uncertainty value of DC Bias Level Accuracy Test (Page 9-39) to
the following information.
DC Bias Level
0 V
1.5 V
2 V
Change 4
Change the Measurement Uncertainty value of 0 m Impedance Measurement Accuracy Test
for 0 m Capacitance Measurement Accuracy Test (Meas. Time Mode: LONG) (Page 9-40) to
the following information.
Test Signal
Standard
Frequency
10 pF
100 kHz
100 pF
1 kHz
100 pF
10 kHz
100 pF
20 kHz
100 pF
100 kHz
1000 pF
100 Hz
1000 pF
1 kHz
1000 pF
100 kHz
0.01 μF
100 Hz
0.01 μF
120 Hz
0.01 μF
1 kHz
0.01 μF
10 kHz
0.01 μF
100 kHz
0.1 μF
1 kHz
0.1 μF
100 kHz
Copyright 2007 Agilent Technologies
C
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Test Limits
±0.002 V
±0.077 V
±0.102 V
Parameter
Measured
Cp
D
Cp
D
Cp
D
2
Cp
D
Cp
D
Cp
D
Cp
D
Cp
D
Cp
D
Cp
D
Cp
D
Cp
D
Cp
D
Cp
D
Cp
D
Measurement
1
Test Result
Uncertainty
V
± 0.088 mV
V
± 0.44 mV
V
± 0.92 mV
Test Limits
Test Result
±0.331 pF
±0.0331
±0.19 pF
±0.0019
±0.44 pF
±0.0044
±1.10 pF
±0.0110
±1.37 pF
±0.0137
±2 pF
±0.0020
±0.0012 nF
±0.0012
±0.0128 nF
±0.0128
± 0.018 nF
±0.0018
± 0.018 nF
±0.0018
± 0.011 nF
±0.0011
± 0.018 nF
±0.0018
± 0.128 nF
±0.0128
± 0.11 nF
±0.0011
± 1.47 nF
±0.0147
Measurement
1
Uncertainty
pF
± 0.0040 pF
± 0.00043
pF
± 0.0074 pF
± 0.000063
pF
± 0.017 pF
± 0.00023
pF
± 0.021 pF
± 0.000091
pF
± 0.019 pF
± 0.00033
pF
± 0.000098 nF
± 0.000063
pF
± 0.000092 nF
± 0.000064
pF
± 0.00012 nF
± 0.00033
pF
± 0.00065 nF
± 0.000022
pF
± 0.00059 nF
± 0.000022
pF
± 0.00076 nF
± 0.000021
pF
± 0.00067 nF
± 0.000036
pF
± 0.00087 nF
± 0.00022
pF
± 0.0068 nF
± 0.000023
pF
± 0.015 nF
± 0.00026