Change 7
Change the Measurement Uncertainty value of 0 m Impedance Measurement Accuracy Test
for 0 m Capacitance Measurement Accuracy Test (DC Bias: ON) (Page 9-41) to the following
information.
Test
Standard
Frequency
1 μF
Change 8
Change the Measurement Uncertainty value of 0 m Impedance Measurement Accuracy Test
for 0 m Resistance Measurement Accuracy Test (Meas. Time Mode: LONG) (Page 9-42) to
the following information.
Test Signal
Standard
Frequency
100 mohm
100 Hz
100 mohm
1 kHz
Change 9
Change the Measurement Uncertainty value of 0 m Impedance Measurement Accuracy Test
for 0 m Resistance Measurement Accuracy Test (Meas. Time Mode: MED) (Page 9-42) to the
following information.
Test Signal
Standard
Frequency
100 mohm
100 Hz
Change 10
Change the Measurement Uncertainty value of 0 m Impedance Measurement Accuracy Test
for 0 m Resistance Measurement Accuracy Test (Meas. Time Mode: SHORT) (Page 9-42) to
the following information.
Test Signal
Standard
Frequency
100 mohm
100 Hz
Copyright 2007 Agilent Technologies
C
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Signal
Parameter
Measured
1 k
Cp
D
Parameter
Measured
R
R
Parameter
Measured
R
Parameter
Measured
R
Test Limits
Test Result
±0.0011 μF
±0.0011
Test Limits
Test Result
±0.52 mohm
mohm
±0.48 mohm
mohm
Test Limits
Test Result
±0.62 mohm
mohm
Test Limits
Test Result
±1.10 mohm
mohm
Measurement
1
Uncertainty
μF
± 0.000066 μF
± 0.000033
Measurement
1
Uncertainty
± 0.030 mohm
± 0.030 mohm
Measurement
1
Uncertainty
± 0.033 mohm
Measurement
1
Uncertainty
± 0.13 mohm