Change 16
Change the Measurement Uncertainty value of 2 m Impedance Measurement Accuracy Test
for 2 m DC Resistance Measurement Accuracy Test (Opt. 001 Only) (Page 9-44) to the
following information.
Standard
100 mohm
100 kohm
Change 17
Change the Measurement Uncertainty value of 4 m Impedance Measurement Accuracy Test
for 4 m Capacitance Measurement Accuracy Test (Page 9-45) to the following information.
Test Signal
Standard
Frequency
100 pF
1000 pF
1 μF
1 μF
Change 18
Change the Measurement Uncertainty value of 4 m Impedance Measurement Accuracy Test
for 4 m Resistance Measurement Accuracy Test (Page 9-45) to the following information.
Test Signal
Standard
Frequency
100 mohm
100 mohm
Copyright 2007 Agilent Technologies
C
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Test Limits
±1.13 mohm
±0.87 kohm
Parameter
Measured
1 kHz
Cp
1 kHz
D
100 Hz
Cp
100 Hz
D
100 Hz
Cp
100 Hz
D
1 Hz
Cp
1 kHz
D
Parameter
Measured
100 Hz
R
1 kHz
R
Measurement
1
Test Result
Uncertainty
mohm
± 0.030 mohm
kohm
± 0.014 kohm
Test Limits
Test Result
±0.19 pF
±0.0019
±0.002 nF
±0.0020
±0.0018 μF
±0.0018
±0.0011 μF
±0.0011
Test Limits
Test Result
±0.84 mohm
mohm
±0.96 mohm
mohm
Measurement
1
Uncertainty
pF
± 0.0078 pF
± 0.000065
pF
± 0.000089 nF
± 0.000062
μF
± 0.000077 μF
± 0.000047
μF
± 0.000071 μF
± 0.000033
Measurement
1
Uncertainty
± 0.030 mohm
± 0.030 mohm