Change 11
Change the Measurement Uncertainty value of 0 m Impedance Measurement Accuracy Test
for 0 m DC Resistance Measurement Accuracy Test (Opt. 001 Only) (Page 9-42) to the
following information.
Standard
100 mohm
100 kohm
Change 12
Change the Measurement Uncertainty value of 1 m Impedance Measurement Accuracy Test
for 1 m Capacitance Measurement Accuracy Test (Page 9-43) to the following information.
Test Signal
Standard
Frequency
100 pF
100 pF
100 pF
20 kHz
100 pF
100 kHz
1000 pF
1 μF
1 μF
1 μF
1 μF
100 kHz
Change 13
Change the Measurement Uncertainty value of 1 m Impedance Measurement Accuracy Test
for 1 m Resistance Measurement Accuracy Test (Page 9-43) to the following information.
Test Signal
Standard
Frequency
100 mohm
100 mohm
Copyright 2007 Agilent Technologies
C
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Test Limits
±0.97 mohm
±0.87 kohm
Parameter
Measured
1 kHz
Cp
D
10 kHz
Cp
D
2
Cp
D
Cp
D
100 Hz
Cp
D
100 Hz
Cp
D
1 kHz
Cp
D
10 kHz
Cp
D
Cp
D
Parameter
Measured
100 Hz
R
1 kHz
R
Measurement
1
Test Result
Uncertainty
mohm
± 0.039 mohm
kohm
± 0.013 kohm
Test Limits
Test Result
±0.19 pF
±0.0019
±0.48 pF
±0.0048
±1.10 pF
±0.0110
±1.37 pF
±0.0137
±0.002 nF
±0.0020
±0.0018 μF
±0.0018
±0.0011 μF
±0.0011
±0.0026 μF
±0.0026
±0.0207 μF
±0.0207
Test Limits
Test Result
±0.60 mohm
mohm
±0.60 mohm
mohm
Measurement
1
Uncertainty
pF
± 0.0077 pF
± 0.000063
pF
± 0.012 pF
± 0.00013
pF
± 0.024 pF
± 0.000085
pF
± 0.010 pF
± 0.00016
pF
± 0.000091 nF
± 0.000063
μF
± 0.000075 μF
± 0.000047
μF
± 0.000069 μF
± 0.000033
μF
± 0.00014 μF
± 0.000061
μF
± 0.00016 μF
± 0.00029
Measurement
1
Uncertainty
± 0.030 mohm
± 0.031 mohm