Ph Sensor Cable Wiring Procedure; Connecting Sensor Cable To Terminal Box - YOKOGAWA PH8ERP User Manual

Kcl refillable type ph sensor
Table of Contents

Advertisement

CAUTION
DO NOT hold the lower part of the sensor when screwing it into a Variopin connecter. Otherwise
the excessive force may damage the sensor.
Fixing Nut
Flow-through Holder
Figure 2.10
How to install the HH350G, PB350G, and PB360G holders, refer to relevant IMs.
2.3

pH Sensor Cable Wiring Procedure

2.3.1

Connecting Sensor Cable to Terminal Box

(1) Open the cable inlet hole in terminal box using the supplied punch tool. The location of
the cable inlet hole is shown by the circle-shaped groove under the case. The end of the
supplied punch tool is put in the center of this circle and it is tapped with appropriate force.
You can punch out the hole along the groove.
pH Sensor
Installing Sensor in Flow-through Holder
< 2. Installation >
O-ring
2-inch Mounting Pipe
F2.10.ai
IM 12B7K1-02E
2-7
11th Edition : Dec.14,2018-00